12
1. Voltage and Current Connections – there are four differential
voltage and four differential current channels, which can be wired
to measure a variety of circuit configurations. Use only Dranetz
voltage leads and current probes for proper operation. Do not
exceed marked ratings.
2. Graphic Color LCD with Touch Screen – the color LCD has an
integral touch screen that is used to select functions for setups,
real time meters, and viewing stored data. Either a clean finger
or a PDA-type stylus can be used and the screen should be
calibrated with whichever you are going to use by selecting
Preferences; Display Preferences, then Calibrate Touch screen
when first turned on. Clean only with soft cloth.
3. Field Replaceable Battery Pack – the internal UPS feature
requires that the internal batteries be properly charged and
maintained. If the battery pack is in need of replacement, de-
energize the connections and power to the unit before opening
the battery door and replacing the pack only with the Dranetz
battery pack PN# BP-PX5. A spare battery pack may be useful
to have.
4. Memory Card slots for data and firmware updates – the Dranetz
supplied Compact Flash memory cards are used for data
storage, as well as to upload new firmware versions when
available. Use only Dranetz supplied memory cards, as the
speed requirements of the instrument can not be satisfied by
many off-the-shelf memory cards. Do not remove the memory
card while monitoring. Only one slot is active at a time.
5. There are three status LEDs and the ON/OFF switch (push for
on, push for off) are located on the bottom of the instrument. AC
power connection for charging the battery and powering the unit
with AC Adapter 117029-G1 is located on the right side of the
instrument. From left to right, the LEDs are:
3 Battery Charge Indicator. LED will light steadily while
battery is fast charging and blink when fully charged.
4 Status Indicator. LED will light steadily when abnormal
condition is detected. The unit is operating normally
when light is off.
5 Power Indicator. LED will blink in a heartbeat fashion
(once per second) when the unit is operating normally.