Connecting Voltage Measurement Cables
Measurement cable set
Description: Voltage measurement cables are provided as
standard accessories and are stored in a cable pouch as part of
the measurement cable set, P/N 116042-G3. Each cable set
consists of a cable and alligator clip.
Voltage Rating: Direct connection of all voltage measurement
cables are rated at 600 Vrms max. For measuring voltages
greater than 600 Vrms, potential transformers (PTs) must be
used.
Optional fused voltage adapter
There are two optional fuse accessory kits available for use with
the measurement cables for PX5. One kit (P/N FVA-1) contains
one fused voltage adapter and one measurement connecting
Red cable 50 cm in length. The other kit (P/N FVA-4) contains
four voltage adapters and four measurement connecting cables
50 cm in length (one Red, one Yellow, one Blue, and one Grey).
WARNING: To avoid the risk of electric shock or burns, always
connect the safety (or earth) ground before making any other
connections.
WARNING: To reduce the risk of fire, electrical shock, or
physical injury it is strongly recommended to fuse the voltage
measurement inputs. Fuses must be located as close to the load
as possible to maximize protection.
WARNING: For continued protection against risk of fire or shock
hazard replace only with same type and rating of recommended
fuse. Use only fast blow type fuse which is rated 600V.
Recommended fuse type is Littelfuse, part number KLKD0.30
rated 600V AC/DC, 0.3A fast blow.
WARNING: Do not replace fuse again if failure is repeated.
Repeated failure indicates a defective condition that will not clear
with replacement of the fuse. Refer condition to a qualified
technician.